メインページはこちらへ
(仕様)
| PERFORMANCE | |
| Range | C1 – 70 to 100% |
| C2 – 0 to 20% | |
| C3 – 0 to 10% | |
| nC4 & iC4 – 0 to 2% | |
| nC5 , iC5, & neoC5 – 0 to 2% | |
| C6+ – 0 to 2% | |
| LOD of Individual Gases | 1000 ppm |
| Response time | T90 < 9 seconds at 3 lpm |
| Flowrate | 0.5 to 3 lpm |
| BTU Accuracy | < 1% of FS |
| Speciation Accuracy | 5% of reading or 0.2% absolute |
| Speciation Repeatability | < 2% of reading or 0.2% absolute |
| ENVIRONMENTAL | |
| Analyzer operational temperature range | 10°C to 35°C |
| Field unit packaged operational temperature range | -40°C to 50°C |
| Housing | General purpose, purged NEMA, explosion proof, or pelican case |
| USER INTERFACE / CONNECTIVITY | Modbus or SOAP over TCP/IP HTML 5 User Interface |
| POWER REQUIREMENTS | 100 to 240 VAC, 50/60 Hz, 200 Watts DC power option available |